Ion Selective Carbon Paste Electrode for Determination of Flavoxate Muscle Relaxant Drug in Pharmaceutical Formulation Potentiometric Determination of Flavoxate
Iranian Journal of Pharmaceutical Sciences,
Vol. 12 Núm. 1 (2016),
15 January 2016
,
Página 45-58
https://doi.org/10.22037/ijps.v12.40790
Resumen
The utility of carbon paste electrode for the determination of flavoxate HCl modified with flavoxate-tetraphenylborate (FLX-TPB) and flavoxate-phosphotungestic acid (FLX-PTA) ion-pairs in batch mode is demonstrated. The electrodes revealed a Nernstian response over a wide concentration ranges 1.39×10-5-1x10-2 mol L-1 and1×10-5-1x10-2 mol L-1 using FLX-TPB and FLX-PTA, respectively. The detection limits of these sensors are 1.39×10-5 mol L-1, and 1x10-6 mol L-1 using FLX-TPB and FLX-PTA, respectively. The best performance was obtained with carbon paste composition of 5% flavoxate-tetraphenylborate or flavoxate-phosphotungestate, 47.5% graphite and 47.5% o- nitro phenyl octyl ether (o-NPOE). The sensors exhibit a very fast response time (5-7 s) and good selectivity in presence of inorganic cations, sugars and aminoacids. The proposed sensors show great improvement in comparison with other previously reported sensors. The sensors were successfully applied to monitoring of flavoxate in pure solution and pharmaceutical formulation (Genurin tablet) with recovery ranges from 97.2 – 101.0% and 98.1-101.6% using FLX-TPB and FLX-PTA, respectively.
- Carbon paste electrode
- Flavoxate
- Potentiometry
Cómo citar
Citas
[2] Sweerman S.C., Martindale, the Complete Drug Reference, the Pharmaceutical Press, London, UK (2005).
[3] El-Gindy A., Abdel-Salam, High-Performance Liquid Chromatographic Determination of Flavoxate Hydrochloride and its Hydrolysis Product, Drug Development and Industrial Pharmacy, (2008) (34): 1311.
[4] British Pharmacopoeia's Stationary Office, British Pharmacopoeia Vol. 1, London, UK (2012) , p. 925.
[5] Zheng Y, Simultaneous Determination Of Ofloxacin And Flavoxate Hydrochloride In Human Plasma By Rp Hplc, Yaowu-Fenxi-Zazhi (1993) (13): 339.
[6] Zarapkar S.S., U.B. Salunkhe, Salunkhe B.B., Doshi,V.J, Sawant, S.V.and Rele R.V., High performance liquid chromatographic determination of 3-methylflavone-8-carboxylic acid, the main active metabolite of flavoxate hydrochloride in human urine, Indian Drugs (1989)( 26): 354.
[7] Sheu M.T., Yeh G.C., Ke H.O., Liquid Chromatographic Determination of Flavoxate HCl in Pharmaceutical Formulation. Chromatogr. B (2001) (751), 79.
[8] Wang Y., Wang T.J., Zhang T.L., Liquid Chromatographic Determination of Flavoxate HCl in Pharmaceutical Formulation Yaowu-Fenxi- Zazhi (2002)( 22): 202.
[9] Zhang C.X., Sun Z.P., Ling D.K, Zheng J.S., Guo J., Li X.Y., Liquid Chromatographic Determination of Flavoxate HCl in Pharmaceutical Formulation, Chromatogr. B (1993)( 123): 287.
[10] Hayashi H., Moffat J. B., Determination of phosphorus and tungsten in heteropoly acids by EDTA-titration, Talanta (1982)( 29): 943.
[11] Waheed A., El-Ries M. A., Carbon Paste- and PVC Membrane Electrodes as Sensitive Sensors for the Determination of Antidiabetic Drugs for Type 2 Diabetic Patients Anal. Sci. (2009)(25): 1431.
[12] Kalcher K., Chemically Modified Carbon Paste Electrodes in Voltammetric AnalysisElectroanalysis, (1990)( 2)V 419.
[13] Kalcher K., Kauffmann J. M., Wang J., Svancara, Vytras I. K., Neuhold C. and Yang Z., ntegration of a Graphite/PMMA Composite Electrode into a Poly(methyl methacrylate) (PMMA) Substrate for Electrochemical Detection in Microchips, Electroanalysis, (1995)( 7): 5.
[14] Svancara I., Vytras K., Barek J. and Zima J., Carbon paste electrodes in modern electroanalysis Crit. Rev. Anal. Chem., (2001)( 31) :311.
[15] Schaller U., Bakker E., Spichiger U. E. and Pretsch E., Ionic additives for ion-selective electrodes based on electrically charged carriers, Anal. Chem., (1994)( 66): 391.
[16] Issa Y.M., Abdel-Fattah H.M., Abdel-Moniem N. B., Chemically Modified Carbon Paste Sensor for Potentiometric Determination of Doxycycline Hydrochloride in Batch and FIA Conditions. Int. Electrochem. Sci., (2013)( 8): 9578.
[17]. Umezawa Y., Umezawa K., Buhlmann P., Hamada Aoki N., H., Nakanishi, M. Sato, K. Ping Xiao and Y. Nishimura, Potentiometric selectivity coefficients of ion-selective electrodes: Part II. Inorganic anions (IUPAC technical report) Pure Appl. Chem., (2002)( 74): 923.
[18]. Ibrahim H., Issa Y. M. and Abu-Shawish H. M. Pharm. Biomed. Anal., (2007)( 44): 8
[19]. Cummings E. A., Mailley P., Linquette-Mailley S., Eggins B. R., McAdams E. T. and McFadden S., Amperometric carbon paste biosensor based on plant tissue for the determination of total flavanol content in beers, Analyst, (1998)( 123): 1975.
[20]. Anchez J. S and Valle M., Determination of anionic surfactants employing potentiometric sensor, Crit. Rev. Anal. Chem., (2005) (35):15-29.
[21] Abu-Shawish H. M., Abu Ghalwa N., Khraish G. I. and J. Hammad, A new potentiometric sensor for determination of pethidine hydrochloride in ampoules and urine, Amer. J. Anal. Chem., (2011) ( 2 ): 56
[22] Zhang X. B., Han Z. X., Z. Fang H., Shen G. L. and Yu R. Q., Electrochemistry of Layer-by-Layer Films: a review Anal. Chim. Acta, (2006) (562): 210
[23]. Buck R. P. and Lindner E., Recommendations for nomenclature of ion selective electrodes (IUPAC Recommendations 1994), Pure Appl. Chem., (1994) (66): 2527.
- Resumen ##plugins.themes.ojsPlusA.frontend.article.viewed##: 98 ##plugins.themes.ojsPlusA.frontend.article.times##
- IJPS_Volume 12_Issue 1_Pages 45-58 (English) ##plugins.themes.ojsPlusA.frontend.article.downloaded##: 24 ##plugins.themes.ojsPlusA.frontend.article.times##